6171L
Description
pH/mV/Temp. bench meter, microprocessor based with dual pH/mV and temperature LCD digital displays, BNC input and ATC probe uses an 8-pin DIN type input. AC/DC Powered.
- Power down memory backup
- Touch keys with audio feedback
- NBS buffer recognition
- 50/60 hz noise rejection
- Electrode offset recognition
- Electrode slope recognition
Specifications
-
Range Resolution Accuracy pH -2.00 to 16.00 pH 0.01pH ± 0.1 % ± 1 digit ORP -1999 to 1999 mV 1.0 mV ± 0.1 % ± 1 digit Temp. 5.0 to 125.0 °C 0.1 °C ± 0.5 °C -
General
- pH buffer recognition model: 6171l - pH 7.00,4.01,10.01
6171lc - pH 6.86,4.0 - pH temp. compensation: Auto/manual -5.0 to 125.0 °C
- pH buffer temp. range: 0 to 60 centigrade
- pH electrode offset recognition: ±90 mV at pH 7.00 or 6.86
- pH electrode slope recognition: ±30 % at pH 4.00,9.18,10.01
- Input impedance: >10exp(12) ohms
- ATC probe: Thermistor (10 kilo_ohm at 25 centigrade)
- Analog output range : standard output /optional ou: Same with the input voltage pH: -900 to 900 mV(pH
- End point sensing & hold: Yes
- Power backup: Memory backup
- Audio feedback: All touch keys
- Display : pH/mV: temp: 12.7 mm high : 12.7 mm high LCD
- Ambient temp. range: 0 to 50 centigrade
- Power : batteries :adaptor: Six AA size batteries : 9 voltage AC adaptor
- Battery life ( alkaline battery ): ~50 hours
- Dimensions (w x l x h ): 250 mm x 240mm x 100mm
- Weight: 1070 gm (batteries included)
- pH buffer recognition model: 6171l - pH 7.00,4.01,10.01
Buy Jenco products from any one of our authorized distributors.