3250

3250

Description

Conductivity/Salinity/Temperature Handheld meter, adjustable cell constants K=0.1, 1.0, 10.0 and selectable 2 or 4 wire cell inputs. Allows programmable temperature co-efficient and reference temperature and includes an optional RS-232C IBM compatible computer interface. Features dual LCD displays for conductivity and temperature and splash proof case. Battery or 115/230 AC powered.
  • Auto/manual ranging
  • CE approved
  • Power down memory backup
  • Splash resistant/watertight case
  • RS-232 output
  • Battery/ac adaptor
  • High 50/60 hz noise rejection
  • Highly reliable mechanical touch keys with tactile & audio feedback
  • User programmable reference temperature & temp. coefficient
  • Selectable 2-wire or 4-wire cells for long cable length & wide measurement range

Specifications

  •  RangeResolutionAccuracy
    Cond.,K=0.10.000 uS/cm to 3.000 mS/cm (K=0.1 over 4 ranges)0.001 uS/cm to 0.001mS/cm, (K=0.1 over 4 ranges)±5% full scale
    Cond.,K=0.10.000 to 9.999 uS/cm0.001 uS/cm±5% full scale
    Cond.,K=0.110.00 to 99.99 uS/cm0.01 uS/cm±5% full scale
    Cond.,K=0.1100.0 to 999.9 uS/cm0.1 uS/cm±5% full scale
    Cond.,K=0.11.000 to 3.000 mS/cm0.001 mS/cm±5% full scale
    Cond.,K=1.00.00 uS/cm to 30.00 mS/cm (K=1.0 over 4 ranges)0.01 uS/cm to 0.01 mS/cm (K=1.0 over 4 ranges)±5% full scale
    Cond.,K=1.00.00 to 99.99 uS/cm0.01 uS/cm±5% full scale
    Cond.,K=1.0100.0 to 999.9 uS/cm0.1 uS/cm±5% full scale
    Cond.,K=1.01.000 to 9.999 mS/cm0.001 mS/cm±5% full scale
    Cond.,K=1.010.00 to 30.00 mS/cm0.01 mS/cm±5% full scale
    Cond.,K=100.0 uS/cm to 300.0 mS/cm (K=10 over 4 ranges)0.1 uS/cm to 0.1 mS/cm (K=10 over 4 ranges)±5% full scale
    Cond.,K=100.0 to 999.9 uS/cm0.1 uS/cm±5% full scale
    Cond.,K=101.000 to 9.999 mS/cm0.001 mS/cm±5% full scale
    Cond.,K=1010.00 to 99.99 mS/cm0.01 mS/cm±5% full scale
    Cond.,K=10100.0 to 300.0 mS/cm0.1 mS/cm±5% full scale
    Salinity0.0 to 80.0 ppt0.1 ppt±1% FS
    Temp.-5.0 to 75.0 Centigrade0.1 Centigrade± 1 Centigrade
  • General

    • Reference temp.: 15.0 to 25.0 °C
    • Temp. coefficient: 0.0% to 40.0%
    • Communication: RS-232
    • Battery life: Greater than 10 hours(continuous operation)
    • Power: Batteries/adaptor, 6 AAA (alkaline)/9V AC adaptor
    • ATC probe: Thermistor, 10 kilo_ohms at 25°C
    • Ambient temp. range: 0 to 50 °C
    • Display :cond/sal: temp.: 15mm high/8.7mm high
    • Dimensions: 222mm x 83mm x 71mm
    • Weight: 450 gm

Share this product